DBL Gas Solutions

AtmoCheck® ONE

AtmoCheck®ONE Modified Atmosphere Analyzer

for checking residual oxygen level in inert gas packaging

A practical hand held, battery operated oxygen analyzer for random sample measurements of residual oxygen levels in Modified Atmosphere / Gas-Flushed food packaging (MAP). This analyzer is ideal for swift, accurate mobile sampling at the packaging machine, in the warehouse or in the laboratory.

Excels with easy operation, short measuring times, and low gas volume sample requirement. The integrated data logger supports traceability and provides complete documentation (HACCP/IFS/ISO).

The software permits fast, uncomplicated communication and is safe and easy to use. It is well equipped for converting handwritten processes to electronic record keeping.

Measuring principle O2 electrochemical measuring cell*
Measuring range 0-100% in 0.1% steps
Measuring time Approx. 10 sec.
Calibration Simple 2-point calibration
Measurement Automatic by needle with integral pump
Measurement memory Circulating memory for 1000 measurements
Interface USB port
Software AtmoCheck® data logger software
Temperature Gas/ambient 40-105°F (5-40°C)
Display Backlit graphic display
Shutdown Automatic when not used for 2 minutes
Housing Shock-resistant plastic
Weight Approx. 15 oz (425g) (without accessories)
Dimensions (HxWxD) 7.4"x4.2"x3.6" (187x106x91mm) (without needle)
Power supply Integrated rechargeable battery (battery charger included in scope of supply)
Battery charger 110-240 VAC
Standards/building regulations company certified to ISO 9001
CE marking as per - EMV 2004/108/EG
- Low voltage directive 2006/95/EG
* estimated service life of O2 sensor in air max. 2 years